DA7200 NIR Analyzer: Instant Water Activity*
The DA7200 measures water activity by shining a bright light on a sample and collecting the reflected spectrum of light. Reflected light from a sample is characteristic of the composition of that sample. In other words, a sample with a water activity of 0.300 will reflect light differently than a sample with a water activity of 0.320.
The DA7200 measures this reflected light and converts it to water activity values using specialized software. Since there is no measurement of the air above the sample, practically no equilibration time is needed. In addition, because the method of detection travels at the speed of light along a fiber-optic cable, the time for a water activity measurement is almost instantaneous.
The DA7200 requires a product specific calibration. Once programmed, the water activity of your product is measured with the touch of a button. No equilibration is necessary.
| Specifications | |
| Warranty: One year parts and labor. |
Accuracy: ± aw |
| Measurement Speed: 6 seconds |
Net Weight: 21 kg (46 lb) |
| Range: 0.100 to 1.000 aw |
Resolution: ± 0.001 aw |
| Dimensions: 0.556 x 0.375 x 0.370 m (with Touch Screen in upright position) |
Reference: Ceramic Automatic |
| Sensors: NR |
Operating Platform: Windows XP, Touch Screen |
| Internal Computer: 1 GHz, 256 MB RAM, 20 GB HD |
Data/Spectral Collection Rate: Approx. 100 spectra/sec |
| Communication Ports: 2xUSB, 1xParallel, 1xLAN, 1xKeyboard, 1xMouse, 1xExternal Monitor |
Power Requirement: 115/230 VAC, 50/60 Hz, 115 W |
| Noise: < 0.00006 AU over 3 sec measuring time |
Wavelength Range: 950-1650 nm |
| Wavelength Accuracy: < 0.3 nm |
Wavelenth Stability: < 0.2 nm/year |
| Ambient Temperature: 5 to 35 °C (41 to 95 °F) |
Optics: F/3 |
| Lamp Lifetime: Approx. 5000 hrs, operator exhangeable |
Grating: Gold-coated ruled stationary grating |
| Sample Dish Capacity: Large sample cut - 154 cm2 (measured 108 cm2) Small sample cup - 44 cm2 (measured 44 cm2 |